Calibrate
The calibration adjusts the internal circuitry to get the best accuracy. Use these functions
to calibrate the vertical and horizontal systems.
For maximum accuracy at any time, run this calibration if the ambient temperature
changes by 5°C or more.
Before running this procedure, do these steps:
1. Disconnect any probes or cables from all channel inputs, otherwise failure or damage
to the scope meter may be occurred.
Utility
Calibrate.
2. Push the
button and select
The calibration screen is shown as in Figure 2-44.
Figure 2-44 The Calibration Screen
NOTE:
30-minutes
The scope meter must have been working or warm-up at least
before
running calibration to get best accuracy.
The scope meter will calibrate parameter of vertical system (CH1, CH2).