1/2
The ALT Trigger menu (Type Pulse)(Page 2/2)
The Alternative menu table (Trigger Type: Pulse Page 2/2)
Menu
Setting
2/2
Setting
Value
HF Reject
ON
OFF
50%
2/2
Term interpretation
■ Auto:
This sweep mode allows the scope meter to acquire waveforms even when it does
not detect a trigger condition. If no trigger condition occurs while the scope meter is
waiting for a specific period (as determined by the time-base setting), it will force itself
to trigger.
When forcing invalid triggers, the scope meter cannot synchronize the waveform,
and the waveform seems to roll across the display. If valid triggers occur, the display
becomes stable on the screen. Any factor results in the un-stability of waveforms can
be detected by Auto Trigger, such as the output of Power supply.
NOTE: When horizontal control is set more than 50 ms/div, Auto mode allows the
oscilloscope to capture without trigger signal.
■ Normal:
The Normal mode allows the scope meter to acquire a waveform only when it is
triggered. If no trigger occurs, the scope meter keeps waiting, and the previous
waveform, if any, will remain on the display.
■ Single:
Go to the next menu page
Figure 2-33 The ALT Trigger menu
Back to the previou menu page
Set the width of the pulse.
Reject high frequency signals or not
Set the trigger level to the center of the signal
Back to the previous menu page
41
Description