AD2817C Operation Manual
Measurement completes
Primary parameter comparison
BIN 1 comparison first;
If fails, BIN 2 comparison;
If fails, BIN 3 comparison
Primary parameter passes some bin
Secondary parameter comparison
Secondary parameter passes
PASS BIN (BIN n)
Figure 4-20 Comparator's comparison and sorting
Please first set limit list before using comparator function. If data in limit table aren't set, it
will be 0.
Perform the following steps to enter limit list setup page:
3. Press SHIFT key at test page, and "SHIFT" at left corner is on.
4. Press LIMIT key to enter setup page, as shown in 4-21.
The following comparator parameters can be set:
AUX:
MODE:
NOM:
BIN1L:
BIN1H:
BIN2L:
BIN2H:
Higher primary parameter (PHI)
Lower primary parameter (PLO)
No pass bins
Secondary parameter fails
Secondary parameter passes (SREJ)
Figure 4-21 Limit list setup page
auxiliary bin on/off;
tolerance mode of primary parameter;
nominal value of primary parameter;
low limit of primary parameter 1;
high limit of primary parameter 1;
low limit of primary parameter 2;
high limit of primary parameter 2;
31
FAIL BIN (OUT)
AUX BIN off
AUX BIN on
AUX BIN