Chapter 6 Handler Interface; General Information - AD Instruments AD2817C Manual De Usuario

Medidor lcr
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

AD2817C Operation Manual

Chapter 6 Handler Interface

6.4 General Information

The handler interface outputs signals to indicate measurement completed, bin sorting
judgments of the comparator function, and Go/No-Go judgments of the list sweep
comparator function. The handler interface also has an input for an external trigger signal.
Using these signals, AD2817C can easily be combined with a component handler and a
system controller to fully automate component testing, sorting, and quality control data
processing to increase production efficiency. Table 9-1 lists the handler interface
specifications.
Specifications
Decision output for
comparator function
Decision output for
list sweep
Output signal
comparator function
(Negative true,
IDX
open collector,
opto-isolated)
EOM
Input
Signal
TRIG
(Opto-isolated)
Table 6-1 Specifications of handler interface
6.5 Signal's Definition
Pin No.
Signal Name
1
/BIN1
2
/BIN2
3
/BIN3
4
NC
5
NC
6
NC
7
NC
Bin number, out of bins, and rejection status
IN/OUT at each sweep point and pass/fail for
the result of sequential comparisons
Analog measurement complete
Full measurement complete.
Display time is not included on the BinNo Disp
page.
Display time is included on the List Sweep
page.
External Trigger, Pulse width≥1 s
Description
Bin sorting judgments.
All bin sorting signal outputs are opto-isolated open
collector.
No Connection
No Connection
No Connection
No Connection