the internally supplied voltage (+5V), or to an externally applied voltage ( EXTV: +5V to
+24V) by setting jumpers.
The electrical characteristics of the DC isolated outputs are divided into two types (See
Table 6-4).
Since the power source for the Comparison Output and Control Output signals are different,
two circuit commons (COM1, COM2) are made available.
Output Signals
Comparison
Signals
/BIN1 - /BIN3
/OUT
/AUX
/PHI
/PLO
/SREJ
Control Signals
/IDX
/EOM
Figure 6-4 DC Isolated output electrical characteristics
Figure 6-5 Position and setup of HANDLER interface's jumpers
Voltage Output Rating
Low
High
≤0.5V
+5V to +24V
+5V to +24V
≤0.5V
Maximum
Current
Internal pull-up voltage:
Circuit common (GND)
External voltage (EXTV1):
6mA
COM1
Internal pull-up voltage:
Circuit common (GND)
6mA
External voltage (EXTV2):
COM2