8
NC
9
NC
10
/OUT
11
/AUX
12
/TRIG
13
/TRIG
14
EXV2
15
EXV2
16
VCC
17
18
19
/PHI
20
/PLO
21
/SREJ
22
NC
23
NC
24
NC
25
NC
26
NC
27
EXTV1
28
EXTV1
29
NC
30
/IDX
31
/EOM
32
COM2
No Connection
No Connection
Bin sorting judgments.
All bin sorting signal outputs are opto-isolated open
collector.
External Trigger:
AD2817A is triggered on the rising edge of a pulse applied
to this pin when the trigger mode is set to EXT mode.
External DC voltage 2:
DC voltage supply pins for DC isolated control signals
(/TRIG, /IDX, /EOM). Setting of internal jumpers must be
changed when using an internal voltage supply.
Internal voltage supply (maximum 100 mA)
Setting of internal jumpers must be changed when using an
internal voltage supply.
Primary Parameter High Reject:
The measurement value is greater than the high limit value
of BIN1–BIN3.
Primary Parameter Low Reject:
The measurement value is less than the low limit value of
BIN1-BIN3.
Secondary Parameter Reject:
The measurement value is not within the range of the
secondary parameter limit.
No Connection
No Connection
No Connection
No Connection
No Connection
External DC voltage 1:
DC voltage supply pins for DC isolated open collector
outputs, /BIN1-/BIN3, /AUX, /OUT, /PHI, /PLO, /SREJ. The
setting of internal jumpers must be changed when using the
internal voltage supply.
No Connection
/IDX signal is asserted when an analog measurement is
completed and AD2817A is ready for the next DUT to be
connected to the UNKNOWN terminals. The measurement
data, however, is not valid until /EOM is asserted.
End of Measurement:
This signal is asserted when the measurement data and
comparison results are valid.
Common for EXTV2