AirSep FreeStyle Manual Del Paciente página 220

Ocultar thumbs Ver también para FreeStyle:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 106
FreeStyle™
Portable and mobile RF communications equipment can effect
medical electrical equipment.
Cannula tubing must be non-kinking, which can be used for a total
length of up to 25 ft (7.6 m) maximum.
Ensure the cannula is fully inserted and secure. This ensures that
the FreeStyle unit can properly detect inspiration for oxygen
delivery.
Storing your FreeStyle for extended periods of time at high
temperatures or with a fully charged or completely discharged
battery can degrade its overall battery life.
Depending upon the temperature of the FreeStyle battery, it can
take several minutes for the charging cycle to start after
connecting to power. This is a normal condition and is intended for
safe charging.
The FreeStyle battery does not need to be fully discharged before
recharging. It is recommended to charge the FreeStyle battery
after each use.
Charging may take several minutes after connecting the power to
initiate, depending on the battery's internal operating temperature.
This is a normal condition and is intended for safe battery
charging. This circumstance is more likely when the battery has
been fully discharged.
®
AirSep
Corporation
MN131-1 rev. B - 02/13
EN
EN-9
Tabla de contenido
loading

Tabla de contenido