calibration (CAL)
There are two ways to calibrate the scale.
The second way is more accurate, but requires the use of multiple standards.
1.
Press the button (CAL) and put the weight that flashes on the scale. The display
stabilizes, you can remove the standard. The scale is calibrated
2.
Press and hold (CoU / aFF) until the scale displays CaL. Place the weight corre-
sponding to the flashing value on the scale.
Standards requirements for calibration
3000g balance:
for 1: 2000g
for 2: 3000 / 2000 / 1000g
2000g balance:
for 1: 1000g
for 2: 2000/ 1500 / 1000 / 500g
200g balance:
for 1: 100g
for 2: : 200 / 150 / 100 / 50g