If the output connector is a 6 pin audio connector. The output pins are:
Pin 2 Input
Pin 3 Output
Pin 5 Signal Ground
It is necessary to jumper pin 1 to pin 6 to enable the RS-232 interface
circuits. Do not connect these pins to the handshaking pins on the interfaced
device.
7.3

OUTPUT FORMAT

The balance will output the weight with the units of measure on one line. The
output is initiated when the [Print] key is pressed or a command is received
over the RS-232 interface.
±w w w • w w w _ _ u u u <cr><lf>
Sign, weight
Units of weight used are:
kg= kilogram, t= tonne, g= gram, lb= Pounds, pcs= parts.
7.4

INPUT COMMANDS FORMAT

The balance can be controlled with the following commands. The commands
must be sent in upper case letters, i.e. "T" not "t". The balance will send the
message "ES" if it does not understand a command that is sent to it.
T<cr><lf>
The uppercase T will tare the balance. This is the same as pressing [Tare].
Z<cr><lf>
The uppercase Z will zero the balance. This is the same as pressing [Zero].
SI<cr><lf>
The SI command will cause the weight to be transmitted over the RS-232
interface. This is the same as pressing the [Print] key.
S0<cr><lf> The S0 (S-zero not S "oh") command will cause the weight to be transmitted
over the RS-232 interface automatically when the results are stable. This is
same as setting the parameter "
S1<cr><lf> The S1 command will cause the weight to be transmitted over the RS-232
interface manually by pressing [Print] or by sending the SI command.