6.12 t1 = AUTOMATIC POWER SWITCH OFF
The scale includes an internal battery. The typical operation time using only
the battery is 50 hours if one load cell is used and about 12 hours if four are
used. The scale includes the "t1" function to turn the power off after 5 minutes
if it is not being used. This function can be disabled if the balance is powered
from the main power supply or if the disruption of power affect the weighing
procedure.
• To set the parameter press the [Func] key to show "t1"
• Press [Print] to see the parameter. Press [Print] to select either
"t1=0" or "t1=1".
"t1= 0" Automatic switch off is enabled.
"t1= 1" Automatic switch off is disabled.
• Press the [Func] key to return to weighing.
6.13 toP = HOLD MAXIMUM VALUE
The display will hold the maximum value weight placed on the pan until the
balance is tared. When a weight is placed on the pan the display will hold the
highest reading until the operator presses [Zero] to reset the display to zero.
When the function is enabled it will show a symbol
• To set the parameter press the [Func] key to show "toP".
• Press [Print] to see the current parameter.
• Press [Print] to select either "toP = 0" or "toP = 1".
"toP = 0"
"toP = 1"
• Press the [Func] key to return to weighing.
6.14 nE = WEIGH IN NEWTONS
The balance can display the unknown weight in Newtons. When Newtons are
selected the weight legend on the display will be turned off and a dash
symbol
will be turned on underneath the 2
• To set the parameter press the [Func] key to show "nE".
• Press [Print] to see the current parameter.
• Press [Print] to select either "nE= 0" or "nE= 1".
"nE= 0"
"nE= 1"
• Press the [Func] key to return to weighing.
© Adam Equipment Company 2005
Hold function is disabled.
Hold function is enabled.
Weigh in kilograms.
Weigh in Newtons.
15
at the second digit.
nd
digit from the left.