Signal Sources
Type of Signal
Signal Increase
Accuracy
Linearity
Frequency Accuracy
Air Conduction Signal Levels
Hz
125
250
Max. dB HL
70
85
Min. dB HL
-10
-10
Masking Signal Levels (NBN)
Hz
125
250
Max. dB HL
70
85
Min. dB HL
-10
-10
Air Conduction Signal Levels
Hz
250
500
Max. dB HL
45
60
Min. dB HL
-10
-10
520-800-MUM · Rev 1.01
DUO
· User's manual
- Pure tones (see frequencies in the Table below).
- Narrow-band noise (NBN). Band as per EN 60645-1.
- Frequency Modulation (FM):
Modulation: sinusoidal – 5Hz. Modulation percentage: 10%.
Continuous and pulsating.
5dB
Sound pressure: +/- 3 dB
Vibratory force: +/- 5 dB
+/- 1 dB
1%
500
750
1000
1500
100
100
100
100
-10
-10
-10
-10
500
750
1000
1500
100
100
100
100
-10
-10
-10
-10
750
1000
1500
2000
60
65
65
65
-10
-10
-10
-10
2000
3000
4000
6000
100
100
100
100
-10
-10
-10
-10
2000
3000
4000
6000
100
100
100
95
-10
-10
-10
-10
3000
4000
6000
65
65
50
-10
-10
-10
8000
85
-10
8000
85
-10
EN 32