SIBELSOUND
To perform an automatic test, select
l
Tone type (depending on the test configuration
selections)
m
Test start (Play)
n
Test stop (Stop)
o
Familiarization
p
Selecting the first ear to examine, Right (R) /
Left (L)
The automatic test is always performed with the autosave by patient's response mode enabled. The procedure is as follows:
•
Press
to create a new patient (see section
•
•
Press
or
to select the first ear to examine.
•
Press
to initiate the familiarization process and instruct the patient on the test procedure. This starts with 1000 Hz tones
at different intensities, which the patient needs to respond to by pressing the Patient's Response Button. This process identifies
a level of intensity close to the threshold, which is used to start the automatic test.
•
After the familiarization process, press
frequencies, following a process similar to the one in the air-conduction tonal test. It will start by issuing a 1000 Hz tone with
an intensity 10 dB above the one identified during the familiarization stage, or with the intensity set as Starting Level on the
automatic test configuration screen.
520-800-MUM · Rev 1.01
DUO
· User's manual
mode on the main screen. The information displayed depends upon:
PATIENT'S DETAILS
to initiate the automatic test. The unit will issue different tones at different
l
m
n
in section 3.6).
EN 27
o
p