7 Conditioning
7.3

Manual cleaning

IMPORTANT: Manual conditioning is possible in exceptional cases if
the national/regional regulations are followed. The national/regional
regulations are to be checked before.
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
NOTICE! Only use Midwest Plus Aerosol Spray.
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
✔ Wear appropriate protective clothing.