5 Operation
32
Operating Instructions Midwest Stylus
5
Operation
NOTICE! Use only burs and diamond polishers that are sharp and
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
CAUTION! A loose or partially removed bur can detach itself from the
head or break off. This may cause injury! Therefore, use the high-speed
handpiece only if the bur is inserted at least 10 mm and is clamped
securely in place.
CAUTION! Insufficient cooling leads to overheating of the
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.
CAUTION! Do not pull the patient's cheek back with the handpiece
while the high-speed handpiece is rotating. This would actuate the
pushbutton, thus creating a risk of burning the patient's oral mucosa.
®
®