ERROR CODES
ERROR
DESCRIPTION
CODES
- -oL - -
Over-range
Err 1
Time Setting Error
Err 2
Date Setting Error
Err 4
Zero Setting Error
Err 6
A/D out of range
Err 9
Check weigh limits
error
FAIL
Calibration error.
© Adam Equipment Company 2015
SUGGESTIONS
Remove weight from the scale.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Enter time using correct format and reasonable values.
Format: hh:mm:ss
Enter date using correct format and reasonable values.
Format: yy:mm:dd
The scale was outside the normal zero setting range either
when it was turned on, or when the [Zero] key was
pressed.
Remove weight from the scale and try re-zeroing again.
Use the [Z/T] key to set the display to zero value.
If the problem persists contact your dealer or Adam
Equipment for assistance.
The values from the A/D converter are outside the normal
range.
Remove the weight from the scale if overloaded.
Make sure the pan is fitted correctly.
Indicates the load cell or the electronics may be faulty.
If the problem persists contact your dealer or Adam
Equipment for assistance.
Shown if the low limit is set higher than the current high
limit. Reset High limit or change the low limit.
Improper calibration (should be within + 10% of the factory
calibration). The old calibration data will be retained until the
calibration process is complete.
If the problem persists contact your dealer or Adam
Equipment for assistance.