ERROR CODES
During the initial power-on testing or during operation, the scale may show an error message. The meaning
of the error messages is described below.
If an error message is shown, repeat the step that caused the message, turning the balance on, carry out the
calibration or other functions. If the error message is still shown contact your dealer for further support.
ERROR CODE
Err 1
Err 2
Err 4
Err 6
© Adam Equipment Company 2013
DESCRIPTION
Time input error.
Date input error
Initial Zero is greater than
allowed (typically 4% of the
maximum
capacity)
power is turned on or when the
[Zero] key is pressed
A/D count is not correct when
turning the scale on.
11
POSSIBLE CAUSES
Tried to set an illegal time, i.e. 26hours
Tried to set an illegal date, i.e. 36
Weight is on the pan when turning the scale
on.
when
Excessive weight on the pan when zeroing the
scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Platform is not installed.
Damaged Load cell.
Damaged Electronics.