Symptom
Mattress appears to
be bottoming out.
Blower Unit Sounds
noisier than usual.
Blower motor inside
pump sounds
excessively noisy
during continuous
use.
Mattress appears to
ignore therapy
settings
Possible Cause
Assess comfort control
Comfort Control
settings.
Settings too soft.
Check patient weight is
Patient Weight.
within limit for system.
Check connections.
Air leak from
manifold.
Check connections in tubeset
Loose hose
connector.
connector.
Check.
Torn or damaged
cell.
Check to ensure the mattress
The blower
is fully inflated.
inside the pump
unit may be
approaching the
end of its life.
Check.
Pump is not
responding to
therapy settings.
Check
16
Action
Adjust as necessary.
Refer to User Manual.
Page 4 Indications.
Remove connection, and
reconnect.
Reconnect if possible,
otherwise call Service
Engineer.
Call service engineer to
replace damaged cell.
Call service engineer to
request a replacement pump
unit (if rental system).
Reset Pump and re-enter
therapy settings.