en - Areas of application
38
Cleaning the filter plate in the water intake
There is a filter plate in the water inlet tube in front of the double
connections for shafted instruments.
This prevents soiling from other disciplines from being deposited in
the lumen of instruments.
The filter plate must be cleaned before every programme.
Unscrew the nut on the clip around the holder for the filter plate and
open the clip.
Pull the right side of the water inlet tube out to the side.
Remove the filter plate.
Remove deposits and soiling from the filter plate with running water,
using a brush if necessary.
Blow the filter plate from the back with compressed air against the
direction of flow.