► Disinfecting Procedures
1. Take out one disinfecting wipe from the package and squeeze out
any excess liquid in order to prevent damage to the meter.
2. Wipe all meter s exterior surface display and buttons. Hold the
meter with the test strip slot pointing down and wipe the area around
the test slot but be careful not to allow excess liquid to get inside.
Keep meter wet with disinfection solution contained in the wipe for a
minimum of 2 minutes for Micro-Kill+™ wipes. Follow the instructions
on the package label of disinfecting wipe.
3. Remove the wipe. Allow the meter surface to dry completely.
4. Discard the used wipes and never reuse them.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact local customer
service at 1-866-469-2632 for assistance.
This device has been validated to withstand up to 10,000 cleaning
and disinfection cycles using the recommended disinfecting wipe/
towelette. The tested number of cycles is estimated by 5 cleaning
and disinfection cycles per day over 5 years, the expected life of the
device. The meter should be replaced after the validated number of