Step 2
Press
to Mark this Test as a Control Solution Test.
With QC displayed, the meter will store your test
result in memory under QC . If you press the
the QC will disappear and this test is no longer a
control solution test.
When doing the control solution test, you have to mark it so that the test
result will not mix with the blood glucose stored in the memory.
Step 3
Apply Control Solution.
Shake the control solution vial thoroughly before
use. Squeeze out a drop and wipe it off, then
squeeze another drop and place it on the tip of
the vial cap.
Hold the meter to move the absorbent hole of
test strip to touch the drop. Once the confirma-
tion window fills completely, the meter will begin
counting down.
To avoid contaminating the control
solution, do not directly apply
control solution onto a strip.
Warning
again,
21