12.1.2 Symbols on the battery charger
In some jurisdictions it is not permissible to dispose of these products with unsor
ted household waste. Disposal that is not in accordance with the regulations of
your country may have a detrimental impact on health and the environment. Please
observe the instructions of your national authority pertaining to return and collec
tion.
Declaration of conformity according to the applicable European directives
Lot number
Protect from moisture
Protection against penetration of solid foreign objects with a diameter greater than
12.5 mm, no protection against water
12.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration sig
nals.
12.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
1 x short
1 x short
12.2.2 Warnings/error signals
Error during use
Beep signal
Vibration signal
None
3 x short
1x before beep signal Battery charger disconnected after start of char
Vibration signal
1x long at interval of
approx. 5 seconds
3 x long
5 x long
Event
Battery charger connected or
battery charger already disconnected prior to
start of charging mode
Charging mode started (3 sec. after connecting
the battery charger)
ging mode
Event
Overheated hydraulic
unit
Battery charge level
under 35%
Battery charge level
under 20%
Required action
Reduce activity.
Charge battery soon.
Charge battery immedi
ately; the product will be
switched off after the next
warning signal.
55