11.1.3 Symbols on the battery charger
Legal manufacturer
Lot number
Declaration of conformity according to the applicable European directives
In some jurisdictions it is not permissible to dispose of these products with unsor
ted household waste. Disposal that is not in accordance with the regulations of
your country may have a detrimental impact on health and the environment. Please
observe the instructions of your national authority pertaining to return and collec
tion.
11.2 Operating states/error signals
The product indicates operating states audibly and visually.
11.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
1 x short
2 x short
1 x long
Switching the Axon-Bus prosthetic system on/off
Charging recept
acle
/
/
11.2.2 Warnings/error signals
Error during use
Charging recept
acle
-
Event
Battery charger connected
Charging begins
Charging ends
Beeper
2 x short
1 x long
Beeper
1x long
-
3 x short
Event
Switch on the Axon-Bus prosthetic system
Switch off the Axon-Bus prosthetic system
Explanation/correction
Error in the Axon-Bus prosthetic system
•
Switch the Axon-Bus prosthetic system off/on
•
Open/close the Axon-Bus prosthetic system
•
Consult prosthetist if error persists
Axon-Bus prosthetic system outside operat
ing temperature range
•
Allow the Axon-Bus prosthetic system to
warm up or cool down (Technical data)
Axon-Bus prosthetic system switches itself
off
•
Charge battery
37