Chapter 12
troublEshootinG
the device can not turn on.
Possible Reason
1. The batteries are drained
or almost drained.
2. The batteries are not inserted properly. 2. Reinstall batteries.
3. The device is broken out.
the device is failure in measuring heart rate.
Possible Reason
1. Electrode and body does not
contact well.
2. Moved when measuring.
3. Electromagnetic interference.
4. Weak signal.
ECG waveform excursion, strong irrelevant waveform or the ECG
waveform displays on the screen, but the prompt info is "Contact?".
Possible Reason
1. Skin is dry or oily.
2. Electrode and body does not.
contact tightly
3. Muscle strain.
ENGLISH
Solution
1. Change batteries.
3. Please contact the local
service center.
Solution
1. Place the electrode correctly.
2. When measuring, please keep
quiet and avoid moving.
3. Keep away from interference source.
4. Change a place (chest) to measure.
Solution
1. Clean skin with soap and water
and eliminate skin crumb and grease
to make the skin wet and greaseless.
2. Press the electrode with
a certain force.
3. Keep hand relax while measuring.