10. Waveform sweeping speed: 20mm/s±10%
11. Input loop current: ≤ 0.1 µA
12. Input impedance  ≥5 MΩ (single ended)
13. Max. input DC offset voltage: ±300mV

10.2 Power supply

battery: 4×AA alkaline batteries (or 4×AA NiMH rechargeable batteries)
external dC input: 5.0V/1.0A;provided AC adapter input voltage:
a.c. 100~240V, 50/60Hz.
auto power-off: timeout setting from 90s to 15 minutes
Threshold of low battery voltage indication: 4.4V +/±0.2V.

10.3 Classification

the type of protection against electric shock:
Internally powered equipment
the degree of protection against electric shock: Type BF applied part
the degree of protection against harmful ingress of liquid:
Ordinary equipment without protection against ingress of liquid.
Electro-magnetic Compatibility: Group I, Class B

10.4 display

LCD type:
Color LCD
Display area: 70.08mm x 52.56mm.

10.5 data memory

Stored data records can be uploaded to the computer by data cable provided
by the manufacturer.

10.6 overall dimension and Weight

Dimension:
(L) 130 mm × (W) 98 mm × (H) 36 mm
Weight:
400g (not including batteries)

10.7 additional declaration

1. Electro-surgical interference suppression: Without the function
of Electro-surgical interference suppression.
2. Defibrillator discharge protection: Without protection against defibrillator
discharge.