lead wires; the "Chest lead" is the external electrodes measurement with 3
lead wires, it is intended for professional use (ie. by doctor). The "limb lead"
or quick measurement by built-in electrodes is recommended for normal
use. The default setting is "limb lead".
• Chest lead: when the measuring mode is set to "chest lead", V lead
is defaulted setting. If you want to obtain specific Chest Lead (V1-V6),
select the lead name (V1-V6) from the pull-down menu.
• Period: "30S", "1min", "5min", "10min", "30min", "continuous"
are optional. The default measuring time is "continuous".
• Hr alarm: "On "and "Off "are optional. The default setting is "On".
• Hr Hi/lo: 35~240bpm, the default value is 120bpm; the low alarm limit
setting range: 30~235bpm, the default value is 60 bpm.
operation:
1. Press the navigation key "/,  or ": Move the cursor
upwards/backwards.
2. Press "" ok key: Conform to option and enter into the corresponding
parameter setting.
3. Press the navigation key " or ": Adjust parameter values.
4. Press "" ok key: Conform the selection, if you do not want to adjust
parameter values, or the setting is accomplished please enter the
" Back" button.

7.1 Creating new id

On the "Option" setup menu, selecting
"New ID" item from the ID pull-down
menu, as shown in figure 7.2.
operation:
1. Press the navigation key "/,  or ":
Move the cursor upwards/backwards.
2. Press the ""ok key: Conform to select
the corresponding numbers or letters.
3. Press the navigation key "/,  or ":
Move the cursor to "Enter" key, press
the OK key "" to accomplish the operation.
During entering the ID number, you can move the cursor to the "Del" button
to correct the numbers or letters input error. When the entering number
exceeds 30 characters, the cursor would move to the "Enter" button
to prompt the user that the entering is finished.
ENGLISH
Figure 7.2