3 REPLACEMENT OF PARTS
Open the case
a
1. For the standing device REF xx-15-xxx, the device
must firstly be removed from the stand tube. To do
this, unscrew the cylinder head screw with internal
hex M8x25 (see a).
2. Remove the basket by loosening the two countersunk
screws with cross recess M5x12 (see b).
3. Tilt the cover slightly (approx. 30°): On one side of the
cover recess, use your finger to press the cover from
back to front until the pin comes out of the receptacle
hole.
4. For table units with REF xx-22-xxx, the device must
firstly be removed from the fixation plate. To do this,
remove the two cap nuts.
b
5. Unscrew the four cylinder head screws with cross
recess M4x10 on the right side (see a) and the four
cylinder head screws on the left side.
6. Now pull the upper part of the case upwards and
lower part downwards.
Replacement of timer
1. Open the case (see "Service Manual, 3 Replacement
of Parts, Open the case").
2. Disconnect electrical connections (see a).
3. Remove the entire timer backwards by loosening the
four crosshead screws (see b).
- 28 / 72 -
b
a
a