Cleaning and disinfecting with SideKick®
For United States only. Confirm availability for your configuration or region. Call Stryker: 1-800-
327- 0770.
Stryker's preferred wipes (2060-000-001 6'' x 10'' or 2060-000-002 9'' x 12'') and spray (2070-000-
001) include the following active ingredients:
n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium
chloride - 0.154%
n-Alkyl (68% C12, 32% C14) dimethyl ethylbenzyl ammonium chloride - 0.154%
Isopropanol - 21.000%
Non-active ingredient: Ethylene Glycol Monobutyl Ether – < 3%
NOTE: For safety information, read the product label.
Clean or disinfect the external product surface with SideKick
To clean:
1. Wipe down the external product surface with a fresh, clean wipe to remove all
visible soils.
2. Repeat as necessary until the external product surface is visibly clean.
3. Wipe dry with a cloth or allow the external product surface to air dry before you
return the product to service.
To disinfect:
1. Clean first.
2. Wipe down the external product surface with a fresh, clean wipe until wet.
3. Allow the external product surface to remain wet for two minutes at room
temperature.
a. To properly disinfect arm caps, ensure the arm caps remain visibly wet with
4. Wipe dry with a cloth or allow the external product surface to air dry before you
return the product to service.
NOTE: Use as many wipes as necessary.
Clean or disinfect the external product surface with SideKick
To clean:
1. Spray SideKick
2. Wipe down the external product surface with a fresh, clean mop, sponge, paper
cloth, or cloth towel to remove all visible soils.
3. Repeat as necessary until the external product surface is visibly clean.
4. Wipe dry with a cloth or allow the external product surface to air dry before you
return the product to service.
To disinfect:
1. Clean first.
2. Spray SideKick
3. Wipe down the external product surface with a fresh, clean mop, sponge, paper
cloth, or cloth towel until wet.
disinfectant for a minimum of 6 minutes.
on a mop, sponge, paper cloth, or cloth towel.
® 
on a mop, sponge, paper cloth, or cloth towel.
® 