AirSep VisionAire Manual Del Usuario página 162

Ocultar thumbs Ver también para VisionAire:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 71
VisionAire™
Problem
Limited or no
oxygen flow.
Condensation
collects in the
oxygen tubing when
you use a
humidifier bottle
Intermittent alarm
sounds at one
second intervals.
MN137-1 rev. H
Probable Cause
Dirty or obstructed
humidifier bottle, or
leak present.
Defective nasal
cannula, or other
oxygen accessories,
i.e. oxygen tubing.
Cannula tubing
loose.
Unit not properly
ventilated.
Elevated operating
temperatures.
See section 2.6
'Safety Features' for
a description of
auditory indicators.
Solution
Remove the humidifier bottle,
and if flow is restored, clean or
replace the humidifier bottle.
Remove and check
accessories for kinks or
obstructions. Replace if
needed.
Check cannula tubing
connection at control panel.
Make sure unit is positioned
away from curtains or drapes,
hot air registers or heaters, Be
certain to place the unit so all
sides are at least 1 foot (30
cm) away from a wall or other
obstruction. Do not place the
unit in a confined area.
Refill humidifier bottle with
COLD water.
DO NOT OVERFILL. Allow
oxygen tubing to dry out, or
replace with new tubing.
Set I/0 power switch to
0 position, use your reserve
oxygen supply (if provided),
and consult your Equipment
Provider immediately.