Domus 4/English
Front panel
2.3.
Pressure Adjust Levels
(1)
Pressure adjust levels controls the air pressure output. When turning up (
output pressure will increase. Vice versa (
pressure output will support the heavier weight patient. In order to find the correct
pressure setting: First inflate the mattress to max pressure level, have the patient lie
on top of the mattress, then decrease the pressure gradually until the patient feel
most comfortable without bottoming out. Have the caregiver insert palm facing up and
place between patient's buttock's area and static cell. Should the palm able to touch
patient's buttock, increase the pressure until the skin is least 1/2" away from the palm.
Always leave at least 1" space between patient's buttock areas to the static cell under
to prevent bottoming out.
Mode
(2)
◼
Max Firm
It will be forced to execute every time the power switch is turned on. This
insures the pump is able to reach its maximum operating pressure. Once the
max pressure level is reached, the pump will automatically switch into
alternating mode. User can also use this function as full mattress inflation while
ingress/egress the patient for better support. There is a 20 min auto-recovery
mode back to alternating cycle to ensure patient always getting best therapy
possible.
Alternate
◼
In Alternate therapy mode, the mattress system will alternate every 10 minutes.
) for decreasing air pressure. Higher-
5
), the