Pressure Set Up
4.3.
Users can adjust pressure of air mattress to a desired softness by adjusting the
comfort keys. Please consult your physician for a suitable setting.
The patient can lie on the mattress after the pump is running in the alternating mode;
this indicates the pressure is reached to the desired softness.
NOTE: Check to see if the suitable pressure is selected by sliding
one hand between the deflated air cells and the patient to feel's
patient's buttock. Users should be able to feel the minimum contact.
Low Pressure Function
4.4.
When an abnormal low pressure is occurred, the Low Pressure indicator (yellow LED)
will light up. Check that connections are correctly made and that they are correctly
installed as per installation instructions.
NOTE: If the pressure level is consistently low, check for any
leakage (tubes or connecting hoses). If necessary, replace any
damaged tubes or hoses. Or contact local qualified dealer for repair.
If the pump unit is equipped with alarm buzzer, the alarm will buzz and the Low
Pressure indicator light up when air pressure is below normal. To mute the alarm,
simply press the
remains until the low-pressure fault condition is solved.
Static Mode
4.5.
Press the
button to suspend the alternating mode of mattress.
Alarm Mute
4.6.
When low-pressure situation occurred, both the LED light and buzzer will sound off
to warn the patient. By pressing the button, it will temporary mute the buzzer so the
care giver may check for possible air leaks. Should the situation not resolved within
3 minutes; the alarm will resume to beep.