9.2.3. A-ZERO
If the Auto-Zero function is active and when no weights are on the pan, the automatic zero tracking
correction is active: the balance remains at a net value equal to zero. The function is useful to cancel small
drift of the zero reading.
9.2.3.1. ZERO OFF
The zero tracking function is inactive.
9.2.3.2. ZERO ON
The function is active and small changes in zero reading are corrected.
9.2.4. FILTER
The setting of the filter menu controls the dynamic response of the balance.
Selections are.
9.2.4.1. SLO
9.2.4.2. FAST
9.2.5. SET DEF
All setting, changed by the setup menu, are valid as soon as they are changed.
However if the balance is powered off they are lost and when powered up the balance will use different
parameters.
To mantain a custom configuration that will be active at power-on the function SET DEF was implemented.
9.2.5.1. STORE
9.2.5.2. END
9.3. CAL E
Once the external calibration function is started the display shows "LOAD" and the reference weight
acquisition is active: a reference weight of E2 class is required for calibration, with value function of the
range, the value of the weight is automatically detected and checked by the balance.
When the value of the reference weight is acquired, the display shows the "UNLOAD" message: at this time
the weight must be removed from the pan and the balance enters the zero acquisition phase.
At the end of the cycle the message "CALIBRATED" appears if the calibration is valid and the new
calibration factor is stored, if anything fails, the message "CAL ERROR" appears and the calibration factor of
the balance remains unchanged.
Notes:
Balances with ranges higher than 100 g use a calibration weight of 100 g and multiples.
Balances with ranges lower than 120 g use a calibration weight of 50 g and multiples.
Page 28 of 45
slow response applicable to disturbed environments (high accuracy weighing)
fast response applicable to quickly changing weights (dosing)
current settings are stored as user default
user default setting is left unchanged
INSRUCTION MANUAL - 60.50.630 9th issue –December 2014
CRYSTAL analytical balance