ENGLISH
9.1. UNIT 2
With this function the balance can be programmed to operate with different measuring units.
CRYSTAL can weight also in ounce (OZ), pounds (LB) and carats (CT).
The balance cal also be programmed for operation as:
STANDARD BALANCE
PIECE-COUNTING BALANCE
PERCENTAGE BALANCE
Select the function and confirm with the ON-OFF key.
Select the desired unit or operating mode, using the MODE key.
Confirm the unit/function selected with the ON-OFF key.
Now the balance is operating with the new unit or function.
9.1.1. STANDARD BALANCE FUNCTION
When not operating as piece-counting or percentage balance the keys act in normal mode.
The key MODE (activated for less than 1 second) switches between gram and the second unit.
If a measuring unit different from gram (main operating unit) is selected, the number of decimal places is
changed and the resolution of the balance (total number of divisions) is reduced, so that the last digit always
exhibits a 1 count increment.
When a second unit is selected the balance automatically stores the information so that, even if the power is
removed, the selection is retained and, when the balance is switched on, the selected unit will be active.
9.1.2. PIECE COUNTING FUNCTION
The function is selected as above with the menu item PCS.
A new reference may be acquired pressing the MODE key for more than 1 second.
The display will show 10 PCS blinking: this is the first sample acquisition quantity.
Pressing the MODE key for less than 1 second, other sample quantities, 20 PCS and 50 PCS, are selected.
Now the reference quantity may be weighted, pressing the ON-OFF key, when the stable symbol lights on
the display, the the new reference value is acquired and the balance starts to display the piece count.
Note:
If the weight of the sample is too small, acquisition is interrupted and the message "PCS ERROR" appears
until the TARE key is activated.
Page 24 of 45
GR ( Gramm )
OZ ( Once )
1 Oz = 28,34952 g
LB ( Libbre )
1 Lb = 453,59237 g
CT ( Carats)
1 Ct = 0,2 g
PCS ( Piece)
%
( Percentage )
INSRUCTION MANUAL - 60.50.630 9th issue –December 2014
CRYSTAL analytical balance