GB/US
12.
Place the filled microcuvette in the cuvette
holder within 40 seconds after filling.
13.
Gently touch the cuvette holder. It will
automatically slide to the measuring
position.
14.
During the measurement "" will be shown.
15.
After approximately 10 seconds, the
hemoglobin value is displayed. The result
will remain on the display as long as the
cuvette holder is in the measuring position.
Do not remeasure the cuvette. When op-
erating on battery power, the analyzer will
automatically turn off after approximately
5 minutes.
16.
Always handle blood specimens with care,
as they might be infectious. Consult
local environmental authorities for proper
disposal.