Airflow test (Figure E)
1. Disconnect the breathing hose.
2. Position the air flow meter on the unit where the breathing hose has been
disconnected.
3. Keep the unit in an upright position.
4. Switch the unit on and set the flow at 160 l/min. The flow amount is satisfactory if
the flow meter float at the indicated level. If it is not, the unit must be inspected (see
Section 4).
6.
Storage
Storing the whole system:
• Remove filters and rechargeable battery.
• Dry the components in a container or cabinet. Store them dry and clean and protect
them from direct sunlight and thermal radiation.
Storing rechargeable batteries:
• Deeply discharged batteries may get damaged after prolonged storage. Charge them
to 50 to 70% prior to storage.
• If storage lasts for over 6 months, charge them in the meantime.
• Do not store rechargeable batteries for prolonged periods outside the recommended
temperature range. This might the remaining capacity and number of potential charge
cycles.
7.
Disposal
This product must not be disposed of as municipal waste. It is therefore marked with the
symbol on the label.
Batteries and rechargeable batteries must not be disposed of as municipal waste. They
are therefore marked with the symbol on the label. Collect batteries and rechargeable
batteries according to local regulations and dispose of at battery collection centers.
20
RSG Safety BV | Powered Air Purifying Respirator T-Air® | June 2021
8.
Technical data
dOverall system
Flow rate of respiratory protective device/helmet/
visor
Flow rate of full face mask
Rated period of service
Operating/storage temperature
Operating/storage area humidity
Noise level
2
International Protection Code
1) For the battery charger and rechargeable batteries, refer to the separate information provided in this chapter.
For other components, refer to the corresponding Instructions for Use.
2) it is mandatory to wear hearing protection (i.e. ear plugs) in combination with all RSG headtops.
Rechargeable battery
Operating/storage temperature
Operating/storage area humidity
Charging temperature
Charging time
Operational life time after a full charge
Rated voltage
Rated capacity
Output power
1) Varies depending on the preset flow rate and the employed filter and facepiece type
Charger
Input voltage
Input current
Output voltage
Output current
International Protection Code
Operating temperature
Operating/storage area humidity
Storage temperature
GB
GB
RSG Safety BV | Powered Air Purifying Respirator T-Air® | June 2021
160/180/200/220 L/min
160 L/min
≥8 hours at all combinations
-10 °C to 50 °C
1)
≤95% relative humidity
1)
depending on air flow, filters and
face piece
IP 65
-10°C to 50 °C
95% relative humidity
0 °C to 50 °C
< 4 hours
≥ 8 hours
1)
14.8 V
5.2 Ah
77 Wh
100~240 VAC
2.0 A(MAX) 50/60 Hz
DC 16.8 V
3.0 A
IP 30
0 °C to 50 °C
≤ 95 % relative humidity
-10 °C to 50 °C
21