Hach GS1440 Instrucciones Para El Usuario página 8

A list of possible errors is shown in
Error
Sensor failure! Contact
technical support.
(or SERVICE SENSOR)
The calibration failed.
(or CAL FAILURE)
Sensor communication lost
(or SENSOR COMM LOST)
Sensor is missing.
(or SENSOR MISSING)
The sensor is not compatible.
(or NOT COMPATIBLE)
4.3 Warning list
A warning does not affect the operation of menus, relays and outputs.
For the SC200 Controller, an exclamation mark shows on the screen. Push the enter key, then
select the instrument. The errors, warnings and reminders for the instrument show on the display.
For the SC1000 Controller, a message window opens. Select CANCEL to save the message to the
message list (or select ENTER to delete the message). To show the message list, select SENSOR
DIAGNOSTIC > MESSAGE LIST.
For the SC4500 Controller, the screen changes to amber. The diagnostics bar shows the error. Push
on the diagnostic bar to show the errors, warnings and reminders.
A list of possible warnings is shown in
Warning
The H₂S gas measurement is too
high.
(or GAS OVERRANGE)
The H₂S gas measurement is
above limit.
(or GAS ABOVE LIMIT)
The H₂S liquid measurement is
too high.
(or LIQUID OVERRANGE)
The H₂S liquid measurement is
above limit.
(or LIQUID ABOVE LIMIT)
Calibration in progress...
(or CAL IN PROGRESS)
Calibration is due
(or CALIBRATION DUE)
8 English
Table
1.
Table 1 Error list
Description
A sensor failure occurred.
The sensor calibration was not
successful.
The gateway is not able to
communicate with the sensor.
The gateway does not
respond to gateway
commands.
The sensor connected to the
gateway is the wrong model.
Table
2.
Table 2 Warning list
Description
The H
S concentration in the air is
2
more than 1000 ppm.
The H
S concentration in the air is
2
more than the H₂S gas limit (or H2S
GAS LIMIT) setting.
The H
S concentration in the liquid
2
is more than 5 mg/L.
The H
S concentration in the liquid
2
is more than the H₂S liquid limit (or
H2S LIQUID LIMIT) setting.
A calibration is in progress.
A calibration is due.
Solution
Contact technical support to repair or replace
the sensor.
Note: If the sensor is without power for 1 year or more,
an internal battery failure can occur.
Refer to the calibration instructions supplied
with the calibration cap and flow regulator.
Make sure that all of the steps shown are
completed as shown in the calibration
instructions.
Examine the sensor cables for damage.
Examine and the sensor cable connections.
Examine the sensor cables for damage.
Examine and the sensor cable connections.
Connect a GS1440 or GS2440EX sensor to
the gateway.
Solution
Move the sensor to a location with a
lower concentration of H
Increase the H₂S gas limit (or H2S
GAS LIMIT) setting as necessary.
Move the sensor to a location with a
lower concentration of H
Increase the H₂S liquid limit (or H2S
LIQUID LIMIT) setting as necessary.
Complete the calibration.
Calibrate the sensor. Refer to the
documentation supplied with the
calibration kit.
S.
2
S.
2
loading

Este manual también es adecuado para:

Gs2440ex