WHAT'S WRONG IF ...

... the entire display does not light up?
... OFF lights up in the display?
... only the upper portions of the display segments light up?
... only the lower portions of the display segments light up?
– the weighing result is unstable?
... the weighing result is obviously incorrect?
12
THEN ...
– the balance is not switched on
– the power cable is not plugged in
– no power reaching instrument
– a momentary disturbance has occured (switch balance off and on again or
disconnect power cable and plug back in)
– the operating voltage selector has been set incorrectly (correct by following
instructions under "Checking the voltage; setting up the balance")
– the microfuse is defective (replace according to instructions under "CARE
AND MAINTENANCE"); if fuse blows again, contact METTLER TOLEDO
Service.
– the power cable has been disconnected and plugged back in or power was
temporarily interrupted (briefly press control bar).
– the weighing range was exceeded
– the calibration weight has been activated (move lever forwards)
– an object was on the weighing pan when the balance was switched on
(remove object, switch balance off/on).
– weighing pan is missing (underload)
– an object was on the weighing pan when the balance was switched on
(remove object, switch balance off/on).
– too much draft (sliding windows closed?)
– balance location is unstable
– integration time has been set too short
– body heat from hand in weighing champer
– object to be weighed is not at room temperature.
– tare has not been pressed
– balance not calibrated or calibrated with an incorrect external weight
– operating voltage switch incorrectly set or line voltage has dropped.