6. Commissioning, operation
6. Commissioning, operation
GB
Push the power button briefly to turn the instrument on. Push the button again to turn it off.
After switching on the firmware version is indicated on the display for approx. 1 second.
6.1 Setup and Configuration
Press the CONFIG key to access the functions of the instrument which are adjustable by
the user. Every time the CONFIG key is pressed, the display switches to the next function.
As soon as a function is configured, press the Enter key to close the Configuration menu,
or the CONFIG key to continue configuration. Configuration and function are set as follows:
Pressure units
The instrument is factory configured to display "PSI". Using the ▲ and ▼ keys (ZERO
and MAX/MIN) you can scroll up and down through the 18 engineering units defined as a
standard, and through the user-defined units or scale. Once the required measuring unit
is displayed, press the Enter key or the CONFIG key. The pressure will then be displayed
in the selected engineering units. A list of available engineering units is given in chapter
"3. Specifications", of this manual. Details on the configuration of user-defined measuring
units is given in chapter "6.2 Supervisor mode".
Automatic Power-off
The function for automatic shut-down can be set in intervals of 1 minute, from 1 to 30
minutes, or it can be turned off (continuous operation). The measuring instrument is deliv-
ered with an automatic power-off after 30 minutes. Define the required interval using the ▲
and ▼ keys. The OFF setting is at the very start of the options before setting for 1 minute.
Displaying battery voltage
The current battery voltage, and the remaining useful life of the battery, are displayed in the
form of a progress bar display.
Indication of the current temperature
The measuring instrument is temperature compensated. This option displays the tempera-
ture measured by the internal sensor. The display value can be switched from degrees
Fahrenheit to degrees Celsius.
Setting damping
Using the ▲ and ▼ keys you can turn this option on and off. With damping enabled, the
measured values from pulsating pressure sources are smoothed.
20
Only connect to test- and calibration installations in the depressurised
condition (atmospheric).