FLAEM NebulFlaem 4.0 Manual Del Usuario página 15

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 73
WARNINGS ON INTERFERENCE RISKS DURING USE IN DIAGNOSTIC IN-
ay
VESTIGATIONS
This device is designed to meet the current requirements for electromagnetic compatibility. As far as EMC
requirements are concerned, electromedical devices require special care when being installed and used,
and are therefore required to be installed and/or used in accordance with the manufacturer's specifications.
Risk of potential electromagnetic interference with other devices. Mobile or portable RF radio and
telecommunication devices (mobile phones or wireless connections) may interfere with the operation of
electromedical devices. The device may be susceptible to electromagnetic interference in the presence of
other devices used for specific diagnosis or treatment. For more information visit www.flaemnuova.it.
ith
CASE HISTORY OF FAULTS AND THEIR RESOLUTION
Before performing any operation, switch off the appliance and disconnect it from the mains.
Problem
re
Thedevice is not
he
working
The device does not
nebulise or nebulises
poorly
ng
The device is louder
than usual
or
If, after checking the conditions described above, the device still is not working properly, we recom-
mend that you contact your trusted dealer or an authorised FLAEM service centre nearest to you. You
can find a list of all Service Centres at http://www.flaemnuova.it/it/info/assistenza
n
ee
Cause
Power cable not correctly
plugged into the appliance
socket or mains power
socket
The medicine was not
inserted into the nebuliser
The nebuliser has not been
assembled correctly.
The nebuliser is obstructed
The accessories are not
properly connected to the
appliance
The pipe is bent or damaged
or twisted
The air filter is dirty
The filter is inserted
Remedy
Plug the power cable correctly into the sockets
Pour the right amount of medicine into the
nebuliser
Disassemble and reassemble the nebuliser
correctly according to the connection diagram
on the cover.
Hygienic preparation of the nebuliser. Deposits
of medicine due to lack of hygienic preparation
of the nebuliser impair its efficiency and
function. Strictly follow the instructions in the
HYGIENIC PREPARATION chapter.
Check the correct connection between the air
inlet of the appliance and the accessories (see
connection diagram on cover).
Unwind the pipe and check it for crushing or
punctures.
Replace it if necessary.
Replace the filter
Insert the filter properly into the housing
15
loading

Este manual también es adecuado para:

PuppyY301p00Ko00p00