2. Pressure measuring range: 0mmHg~300mmHg
3. Cuff inflation time: <20 seconds (typical adult cuff)
4. Accuracy of pressure measurement: ±3mmHg
5. Overpressure protection limit
Adult: ≤300mmHg (39.9kPa)
6. Blood pressure measuring range:
SYS: 60mmHg~240mmHg
7. Blood pressure measuring accuracy: Maximal mean difference: ±5mmHg
8. Maximal standard deviation: 8mmHg
4.2 SpO
Measurement
2
1. Technique: optical with dual-wavelength
LED wavelength: Red light: 663nm, Infrared light: 890nm
Maximal optical output power: less than 2mW maximum average
2. SpO
measuring range: 35%~100%
2
3. SpO
measuring accuracy: Arms is not greater than 3% for SpO
2
Note: Arms is defined as root-mean-square value of deviation according to ISO 9919
4. Low perfusion performance: the above declaration is still attained while the amplitude
modulation ratio is as low as 0.6%.

4.3 Pulse Rate Measurement

1. PR measuring range: 30bpm~240bpm
2. Pulse rate measuring accuracy: ±2bpm or ±2%, which is greater
ENGLISH
DIA: 30mmHg~180mmHg
range from 70% to 100%
2
92