4.4 Temperature Measurement

1. Measuring range: 32.0°C~43.0°C
2. Measuring accuracy: ±0.2°C is for TEMP range from 35.0°C to 42.0°C, and ±0.3°C
is for the rest.
3. Response time: ≤5s

4.5 Blood Glucose Measurement (Optional)

1. Technique: Amperometric, glucose oxidase
2. Measuring range: 1.1mmol/L~33.3mmol/L (20~600mg/dL)
3. Minimum sample size: 1ul.
4. Measuring time: 10 seconds

4.6 ECG Measurement (Optional)

1. Heart Rate measuring range: 30bpm~240bpm
2. Heart Rate measuring accuracy: ±2bpm or ±2% whichever is greater
3. Display scale: 5.0mm/mV±10%
4. Common-mode rejection ratio (CMRR): ≥60dB

4.7 Others

4.7.1 Operating Environment

1. Operating temperature: 5°C~40°C; Relative humidity: 15%~93%;
Atmospheric pressure: 70.0kPa~106.0kPa; Power supply: a.c. 110V-240V AC, 50/60Hz;