Sensitivity Adjustment
4-1. Automatic Sensitivity Adjustment (GX-L Series Only)
This function automatically adjusts the sensitivity of the balance according to ambient temperature
change, set time or interval time using the internal mass. For details, refer to "7-1. Automatic Sensitivity
Adjustment" in the separate manual.
4-2.
Sensitivity Adjustment Using the Internal Mass (GX-L Series Only)
Sensitivity adjustment using the internal mass can be performed with one key press.
1. Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing pan.
2. Press the CAL key. The balance displays
3. The sensitivity of the balance is adjusted using the internal mass. Do not allow vibration or drafts to affect
the balance.
4. After sensitivity adjustment, if GLP output is set, a "sensitivity adjustment report" is output or stored to
data memory.
5. The balance returns to weighing mode automatically.
About the internal mass
The value of the internal mass may change due to factors such as the operating environment and aging.
Correct the internal mass value as necessary. Refer to "7-6 Correcting the Internal Mass Value (GX-L Series
Only)" in the separate manual.
Since the internal mass is about 1 kg, the possibility of error may increase as the weighing value increases.
To maintain the weighing accuracy, perform the sensitivity adjustment using an external weight periodically,
as described in
"4-3. Sensitivity Adjustment Using an External
This indicates that the balance is importing calibration data. Do not allow
vibration or drafts to affect the balance while this indicator is displayed.
After calibration, the balance returns to indicate the previous display.
.
Weight".