4.4
TDS / salinity measurement
To perform a TDS / salinity measurement, follow the same procedure as for a conductivity measurement.
Press Mode to switch between conductivity, TDS and salinity measurement mode.
4.5
Setup
4.5.1
Set temperature correction coefficient
Press the Setup key (FE30) or press and hold the Mode/Setup key (FG3) until the setup icon appears
on the display and the current standard blinks. Press Read to ignore standard setting; the meter changes
to the α-coefficient setting automatically.
Use or to increase or decrease the temperature correction value. Press Read to confirm your setting.
Continue with reference temperature setting or press Exit to go back to the measurement screen.
4.5.2
Set reference temperature
After confirming the temperature correction coefficient, the reference temperature appears on the display.
Use or to toggle between 25 °C and 20 °C. Press Read to confirm your selection. Continue with
temperature unit setting or press Exit to go back to the measurement screen.
4.5.3
Set temperature unit
After selecting the reference temperature, the current temperature unit blinks. Use or to toggle
between °C and °F. Press Read to confirm your selection. Continue with TDS factor setting or press Exit
to go back to the measurement screen.
4.5.4
Set TDS factor
After selecting the temperature unit, the current TDS factor blinks. Use or to increase or decrease
the value. Press Read to confirm your setting. The meter will exit automatically to the measurement
screen.
METTLER TOLEDO FiveEasy™ / FiveGo™ conductivity meter
Five_cond.indb 9
Operating the conductivity meter
13.03.2007 16:07:41
9