4 Tap Start verification.
5 Enter the lot number if Methods > Verification > Check buffer lot number is active and confirm with OK.
The verification point, sensor name, sensor status and the endpoint type are displayed.
6 Tap Read.
7 The measurement stops when the endpoint criterion is fulfilled. See [Endpoint criterion Page 16].
A window pops up with verification data.
8 Tap OK.
5.5 Starting a measurement
1 Select an appropriate method according to [Selecting a method Page 14].
2 Select a sensor if necessary according to [Selecting a sensor Page 14].
3 Tap Sample ID to enter the Sample ID if necessary.
4 Place the sensor in the sample.
5 Tap Read.
6 The measurement stops when the endpoint criterion is fulfilled. See [Endpoint criterion Page 16].
7 Tap Confirm if Menu > Settings > Analysis settings > Confirm end of analysis is active.
5.6 Interrupting analysis
Ongoing analyses can be interrupted anytime.
To interrupt an analysis, proceed as follows:
1 Tap Terminate.
2 Confirm with OK to go back to Home screen.
- or -
Tap Cancel to continue the analysis.
5.7 Endpoint criterion
The measurement will stop based on the setting in Methods > Measure > Endpoint type.
If Endpoint type is defined as:
Automatic: The measurement will stop automatically, based on the programmed stability criteria.
Manual: The measurement will only stop when you tap Manual endpoint.
Timed: The measurement will stop after the defined time in Methods > Measure > Endpoint time.
If Home screen > Menu > Settings > Analysis settings > Allow taking of manual endpoint is active, you can
also stop the measurement manually by tapping Manual endpoint during the measurement, whatever the
Endpoint type is defined for.
5.8 Measurement status
Measurement Reading
Measurement
limit
Inactive
Active
16
In progress
Within limit
Out of limit
font: light blue
background: white
font: light blue
font: red
background: white
Endpoint reached
Within limit
Out of limit
font: white
background: dark blue background: light
blue
font: white
background: green
background: red
SevenDirect™ SD20