Message
Data memory error
No matching data found in memory
Sensor ID already exists, previous SN
will be overwritten
Update failed
Export failed
8.5 E rror limits
Message
Out of range, determine again
Buffer/standard temp. out of range
Offset out of range
Slope out of range
Wrong buffer
Invalid pH for user-defined buffer
ATC measured temperature is different
to the user-defined value
Description and Resolution
The data could not be stored into memory.
●
Switch the meter off and back on.
●
Contact METTLER TOLEDO service if the problem per-
sists.
The entered filter criterion does not exist.
●
Enter a new filter criterion.
Two sensors with the same ID but different SN are not al-
lowed in the meter. If a different SN has been entered for
this sensor ID previously, the old SN will be over w rit t en.
●
Enter a different Sensor ID in order to keep the pre v i-
ous ID and SN.
The software update process failed. This could be due to
the following reasons:
●
The USB stick is not connected or it is disconnected
dur i ng the update process
●
The update software is not in the correct folder
The exporting process failed. This could be due to the fol-
low i ng reasons:
●
The USB stick is not connected or it is disconnected
dur i ng the exporting process
●
The USB stick is full
Range not accepted
pH
mV
T (pH)
| Eref1-Eb | > 60 mV
Slope < 85% or > 110%
I ∆Eref1I < 10 mV
I ∆pHI < 1 pH
| T(ATC)T(buffer) | > 1 °C
<-2.000 or > 20.000
<-2000.0 or > 2000.0
< 5 or > 50 °C
Maintenance