Error Limits - Mettler Toledo SevenCompact S220 Instrucciones De Manejo

Ocultar thumbs Ver también para SevenCompact S220:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Message
Data memory error
No matching data found in memory
Sensor ID already exists, previous SN
will be overwritten
Update failed
Export failed
8.5  E rror limits
Message
Out of range, determine again
Buffer/standard temp. out of range
Offset out of range
Slope out of range
Wrong buffer
Invalid pH for user-defined buffer
ATC measured temperature is different
to the user-defined value
Description and Resolution
The data could not be stored into memory.
Switch the meter off and back on.
Contact METTLER TOLEDO service if the problem per-
sists.
The entered filter criterion does not exist.
Enter a new filter criterion.
Two sensors with the same ID but different SN are not al-
lowed in the meter. If a different SN has been entered for
this sensor ID previously, the old SN will be over w rit t en.
Enter a different Sensor ID in order to keep the pre v i-
ous ID and SN.
The software update process failed. This could be due to
the following reasons:
The USB stick is not connected or it is disconnected
dur i ng the update process
The update software is not in the correct folder
The exporting process failed. This could be due to the fol-
low i ng reasons:
The USB stick is not connected or it is disconnected
dur i ng the exporting process
The USB stick is full
Range not accepted
pH
mV
T (pH)
| Eref1-Eb | > 60 mV
Slope < 85% or > 110%
I ∆Eref1I < 10 mV
I ∆pHI < 1 pH
| T(ATC)T(buffer) | > 1 °C
<-2.000 or > 20.000
<-2000.0 or > 2000.0
< 5 or > 50 °C