To self-test for any of the above, perform the following steps:
1. Place new batteries in the device.
2. Verify the device is off.
3. Insert one new leadwire into two new electrodes.
4. Place the new electrodes on your forearm as shown in the figure below.
5. Insert the leadwire in Channel 1.
6. Turn your device on.
7. Select PPR 5. This is a continuous treatment program. The lights on the device should be
steady.
8. Slowly increase the amplitude (intensity) until you can feel it. If you do not get any
sensation, lower the amplitude (intensity) to zero and rotate the leadwire 90 degrees. Slowly
increase the amplitude (intensity).
NO SENSATION = Call the repair department
SENSATION IS FELT - EVEN IF WEAK = device is working properly. You may need to reposition
the electrodes or contact your clinician.
9. Repeat Steps 1 through 8 for Channel 2.