3.2.1.5 PPR 5 — Summary
PPR 5 offers continuous stimulation unless used with an external hand or foot
remote switch (see 3.1.1 for details). As such, it is intended for use as a gait training
protocol. The intensity of each channel is independently adjustable. For the remote
hand switch, Channel 1 is active when the trigger switch is depressed, and Channel
2 is active when the trigger switch is released. For the foot switch, Channel 1 is
active when the weight of the foot is released from the foot switch, and Channel 2 is
active when the weight of the foot is applied to the foot switch.
Details
ADJUSTABLE PARAMETERS
Waveform
Treatment Time
PRESET PARAMETERS
Waveform
Treatment Time (Timer)
Pulse Rate
Pulse Duration (Width)
3.2.1.6 PPR 6 — Summary
This regimen is a treatment for reduction of muscle spasms with a 2:1 ON/OFF
ratio.
Details
ADJUSTABLE PARAMETERS
Waveform
Treatment Time
Off Time
Pulse Rate
Channel Cycling
PRESET PARAMETERS
Waveform
Treatment Time
Channel 1 Ramp-up time
Channel 1 ON time
Channel 1 Ramp-down time
Channel 2 Ramp-up time
Channel 2 ON time
Channel 2 Ramp-down time
RANGE
Symmetric or Asymmetric
Unlimited, or 5-60 minutes
DEFAULT SETTINGS
Symmetric
30 minutes
35 Hz
300 s
RANGE
Symmetric or Asymmetric
Unlimited, or 5-60 minutes
0 to 60 seconds
2 to 150 Hz
Synchronous or Asynchronous
DEFAULT SETTINGS
Symmetric
30 minutes
2 seconds
10 seconds
2 seconds
2 seconds
10 seconds
2 seconds