Press the [ENTER] key to set the zero point (the display of UNIT
•
WEIGHT will show and flash for a few seconds, then change to "XXXX"
to set the weight of the weight to be calibrated).
•
Press the [CE] button and use the numeric buttons to set the weight
of the weight to be calibrated, e.g. if it is 10 kg, set it with the [1] + [0]
buttons.
Place the weight on the large pan - its weight will be shown on the
•
UNIT WEIGHT display, i.e. first it will flash "XX" and then it will change
to the displayed weight according to the pattern "XX.XXX kg". After
that, the calibration will be completed and the scale must be restarted.
After restarting, check the correctness of weighing on the weight used
•
for calibration - if the indicated weight is correct, the calibration was
successful. Otherwise, the scale must be re-calibrated.
For additional (external) weight - optional
IMPORTANT: to connect an additional scale to the D-SUB type connector, you
need a cable with a DB9 or RS232 plug (not included).
Make sure that nothing is placed on the attached external scale / pan.
•
Turn on the device with the ON / OFF button (G) located on the back
•
of the casing (F), and then press the numeric buttons [1], [3], [1], [3]
on the keypad and while the scale is self-checking, press the button
[ENTER] - WEIGHT will show "SCALE3", UNIT WEIGHT "CAP XXXX" and
TOTAL "d = XX".
•
Press the [UNIT] button to select the weight unit for calibration - kg or
lb.
•
Press the [ENTER] key to set the zero point (the display of UNIT
WEIGHT will show and flash for a few seconds, then change to "XXXX"
to set the weight of the weight to be calibrated).
Press the [CE] button and use the numeric buttons to set the weight
•
of the weight to be calibrated, e.g. if it is 100 kg, set it with the [1] +
[0] + [0] buttons.
•
Place the weight on the large pan - its weight will be shown on the
UNIT WEIGHT display, i.e. first it will flash "XX" and then it will change
to the displayed weight according to the pattern "XX.XXX kg". After
that, the calibration will be completed and the scale must be restarted.
After restarting, check the correctness of weighing on the weight used
•
for calibration - if the indicated weight is correct, the calibration was
successful. Otherwise, the scale must be re-calibrated.
3.3.
Working with the device
Display