PRECAUTIONS
Before connecting the unit ensure that the power
supply is 230 V, 50/60 Hz and earthed. It should also
be protected by a thermal magnetic cut-out. Consult
your installer concerning these matters.
Do not allow children or unskilled personnel to
handle the unit.
Use long gauntlets and muffle tongs whenever you
put muffles into the tank or take them out.
Open the cover carefully: steam pouring out from the
tank could cause burns.
Always keep the unit's cover closed while
polymerisation programs are running.
TECHNICAL DATA
Height x width x depth:
Tank dimensions:
Weight:
Power supply:
Consumption:
Nº muffles with flange:
Ref. 080405
420 x 270 x 480 mm
335 x 180 x 350 mm
15 kg
230 V, 50/60 Hz
1750 W
9
Never place more than 9 muffles on the polymerising
tray.
Always keep at least 20 cm of water in the bottom of
the tank, as the heating elements may be damaged if
the level is lower.
Always leave a gap of at least 15 cm between the
unit's ventilation grille and the nearest wall or other
object.
Before plugging in the unit, fill the tank with water to
10 cm below the cover.
TALLERES MESTRAITUA, S.L. recommends
polymerising at around 95 ºC. At higher temperatures
too much steam may be generated.
Ref. 080407
510 x 420 x 500 mm
350 x 320 x 400 mm
20 kg
230 V, 50/60 Hz
3000 W
18
English
9