Result Number
Message
ANALYZER
EXPIRED
21 / 22
NO DAYS
REMAIN
ANALYZER
EXPIRED
23
NO TESTS
REMAIN
INVALID BARCODE
ON TEST DEVICE
24
REPEAT TEST WITH
NEW TEST DEVICE
VERIFY FAIL
CONSULT
25
MANUAL
ERROR 25
RESULT INVALID
FLU A: + FLU B: +
26
CONFIRM RESULT
ERROR 26
ANALYZER
FAILURE
27
CALL BD
ERROR 27
ANALYZER
FAILURE
28
CALL BD
ERROR 28
ANALYZER
FAILURE
29
CALL BD
ERROR 29
Possible Cause(s)
The shelf life of the
BD Veritor Plus Analyzer has
expired. The maximum shelf-
life of the BD Veritor Plus
Analyzer from the date of
manufacture is 34 months
(regardless of time of first
use).
or
The operational life of the
BD Veritor Plus Analyzer
has expired. The maximum
number of months from first
use is 24 months.
The operational life of the
BD Veritor Plus Analyzer has
expired. The maximum
number of tests performed
by a BD Veritor Plus
Analyzer is 3500 runs.
The test device barcode
cannot be read correctly.
The result of a BD Veritor
System Verification Cartridge
run is unsuccessful.
The BD Veritor Plus Analyzer
has detected a Flu A and Flu
B co-infection.
A hardware anomaly has
occurred.
A hardware anomaly has
occurred.
A hardware anomaly has
occurred.
Corrective Action(s)
Contact your distributor to order a new BD Veritor Plus
Analyzer.
Contact your distributor to order a new BD Veritor Plus
Analyzer.
Repeat the assay. On subsequent occurrences, contact
BD.
Ensure that no foreign matter is present on the BD Veritor
System Verification Cartridge assay area, and repeat the
run. On subsequent occurrences, contact BD.
The BD Veritor Plus Analyzer reports dual positive
influenza A and influenza B results as "Result Invalid."
Specimens generating a "Result Invalid" should be
retested. Upon retesting, if the specimen produces a
"Result Invalid" the user may want to consider other
methods to determine whether the sample is positive or
negative for influenza virus.
Contact BD.
Contact BD.
Contact BD.
25