BD Veritor™ Plus Analyzer Instructions for Use
Result Number
11
12
13
14
15
16
17
18
BACKGROUND INVALID
19
20
24
Message
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 11
NEGATIVE CONTROL
Cross-reacting material
LINE INVALID
present.
CONSULT MANUAL
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 13
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 14
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 15
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 16
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 17
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 18
The assay background area
CONSULT MANUAL
is outside of the normal
ERROR 19
range.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 20
Possible Cause(s)
Corrective Action(s)
Contact BD.
On first occurrence, repeat the assay. On subsequent
occurrences, contact BD.
Contact BD.
Contact BD.
Contact BD.
Contact BD.
Contact BD.
Contact BD.
Remove the device and cycle the instrument power.
• If the instrument displays an error during power-on,
contact BD.
• If no error message appears, obtain an additional
sample and repeat the assay If the issue recurs, contact
BD.
Contact BD.